FDA: Transfer of a Premarket Notification (510(k)) Clearance – Questions and Answers

1. The 510(k) database will not show the new holder

  • New in final (Q2, footnote 17): Once cleared, the holder is listed as the "applicant" in the 510(k) Premarket Notification Database, and the applicant name does not change even if the 510(k) is transferred.
  • Combined with Q11 (FDA is not required to update records on request), this means the registration and listing system, not the 510(k) database, is how FDA identifies the current holder.
  • Action: Acquirers should not expect the public 510(k) database to reflect a transfer. Keep transfer documentation (purchase agreements, assignment letters) available to demonstrate holder status to customers, notified bodies, and inspectors.

2. A new listing is explicitly required to reflect a transfer

  • New in final (Q5, Q8): When a transfer occurs, the new holder must create a new listing to reflect the transfer if introducing the device into commercial distribution (21 CFR 807.28(b)(2)). The system assigns a new unique listing number when the new holder creates the listing.
  • New in final (Q5, footnote 22): When a new listing number is created, labelers must also update the GUDID record with the new listing number (21 CFR 830.330(b)).
  • Action: Build both steps into the transfer project plan: new listing in FDA's registration and listing system, followed by GUDID record updates carrying the new listing number.

3. Which premarket submission number to list is now spelled out (new Q7)

  • Unchanged device: the new holder must enter the original FDA-assigned premarket submission number, unless it submits a new 510(k), in which case it uses the new number.
  • Significantly changed device (design, components, method of manufacture, or intended use): the new holder must generally submit a new 510(k) and, after clearance, list under the new number.
  • The draft covered this only briefly in the Background section; the final gives it a dedicated question.
  • Action: Before listing, run any post-acquisition changes (e.g., manufacturing site moves, supplier or component changes) through the "Deciding When to Submit a 510(k)" guidances to confirm which number applies.

4. Firmer position on a single 510(k) holder, with rationale (Q8)

  • Draft: FDA "believes" there can be only one holder at a time.
  • Final: a flat "No" to simultaneous holders, with stated reasons: compliance with legal requirements, tracking of signals and adverse events, and communication between sponsors and FDA.
  • The definition of "510(k) holder" now says "the singular person" who possesses the clearance, and notes the holder may be a manufacturer, which may be a specifications developer.
  • New reassurances and clarifications:
  • The guidance is not intended to limit a person's ability to transfer a 510(k) or enter business arrangements; FDA acknowledges complicated manufacturing and distribution arrangements.
  • Footnote 34: FD&C Act requirements may apply regardless of whether someone is the 510(k) holder.
  • Case-specific questions on designating the holder can go to the 510(k) Staff (301-796-5640).
  • Action: Co-ownership, licensing, or split-rights deals should designate one holder clearly in the contract; other parties (contract manufacturers, sterilizers, repackagers, relabelers) list under the holder's number per Q9.

5. Obligations of the previous holder are clearer

  • The final states the guidance applies to both the previous and the new holder (Section II, Scope).
  • A previous holder that stops performing an activity identified on the listing (e.g., manufacturing) must update the listing (21 CFR 807.28(a)).
  • Discontinuation duty is now framed as applying to "a 510(k) holder who discontinues commercial distribution" (draft said "an owner or operator").
  • Action: Sellers should include listing updates and discontinuations in their divestiture close-out checklist, not just the buyer's onboarding plan.

6. Broader description of who must submit a 510(k) (Q2)

  • Draft: "You are required..." Final: "You generally are required... and you are a manufacturer, including a:"
  • The list was restructured:
  • "Domestic manufacturer" and the separate "Foreign manufacturer" items are merged into "Finished device manufacturer" (a footnote confirms a manufacturer can be domestic or foreign).
  • New separate item: manufacturers of components or accessories ready for health-related use and packaged or labeled for that purpose, or devices that must be further processed by a licensed practitioner or other qualified person to meet a particular patient's needs (21 CFR 807.20(a)(6)).
  • "Specification developer" is renamed "specifications developer" (matching the regulation).
  • The Introduction now also states that a 510(k) is required for significant changes to a device already in distribution (807.81(a)(3)), not only first introductions.

7. Rationale for "no new 510(k)" is stated more definitively (Q4)

  • Draft: the new holder does not need a new 510(k) "if the new 510(k) holder would not be proposing" to begin introduction into commerce.
  • Final: no new 510(k) is needed "since the new 510(k) holder would not be proposing" to begin introduction.
  • This removes a conditional that some readers saw as an extra test. The real condition remains that the device is not significantly changed or modified.

8. UDI/GUDID timing made explicit (Q10)

  • Footnote 42 now quotes 21 CFR 830.330(b): updated GUDID information must be submitted no later than the date a device is first labeled with the changed information, or within 10 business days of the change if the information does not appear on the label.
  • UDI wording now says the label of every device and every device package must bear a UDI.
  • Labeling updates (manufacturer, packer, or distributor name and place of business under 21 CFR 801.1(a)) remain flagged as potentially required.
  • Action: Align label changeover and GUDID updates so the labeler name change is submitted on or before first shipment under new labels.

9. Restructured as a full Q&A (5 questions → 12)

Substantive content from the draft's Background section was moved into, and expanded across, stand-alone questions. New or split-out questions:

  • Q1: Who needs to register and list?
  • Q3: When does a manufacturer need to submit a 510(k)?
  • Q5: What are the listing requirements for a new 510(k) holder?
  • Q6: When should a new 510(k) holder register and list?
  • Q7: Which cleared 510(k) number must the new holder enter?
  • Q10: Is there other information a new holder may need to submit?
  • Q11: Who must update information about a transfer or sale?
  • Q12: How should I obtain copies of 510(k) records?

Q6 also adds that listing information may be changed at other times (e.g., when a device is introduced, changed, or removed from distribution) under 21 CFR 807.22(b)(4), in addition to the October 1 – December 31 annual window.

10. Definitions updated

  • New: "510(k) clearance" defined as FDA's order declaring the device substantially equivalent to a legally marketed predicate (21 CFR 807.100(a)(1)).
  • Contract manufacturer broadened from making a "finished device" to making "a device" for another person, which may pull component or subassembly contract manufacturers into scope.
  • Remanufacturer citation updated from 21 CFR 820.3(w) to 820.3(a), reflecting the amended Part 820 (QMSR).
  • Specifications developer now also cites 21 CFR 807.20(a)(1).
  • Definitions are now consistently framed as "a person who..."