1. Introduction, Background, Scope, and Templates
Scope is now tied to the 564 declaration rather than pathogen novelty. The draft covered tests to detect a "newly identified, previously unknown, or unusual pathogen(s)," or a known pathogen in a newly identified or unusual clinical presentation. The final covers tests intended to detect a pathogen to aid diagnosis of a serious or life-threatening disease or condition *for which an applicable 564 declaration has been issued*. The introduction similarly now references a determination under section 564(b)(1).
Home collection kits and certain software are explicitly in scope. A new footnote states the guidance includes recommendations on home collection kits and certain software used with diagnostic tests.
New flexibility language for special situations. The final adds that outbreaks inherently carry uncertainty, that recommended approaches should be adjusted for certain situations, and gives "matters of national security" as an example.
De-emphasis of COVID-19 as the reference model. The draft stated most examples were based on SARS-CoV-2 and similar respiratory viruses; that statement was removed, and the final states the recommendations are intended to be broadly applicable. Many COVID-specific examples throughout the document were also removed. The mpox and COVID-19 EUA templates are still referenced, but now framed as examples of "approaches taken during prior emergencies."
Background trimmed. The list of eight prior IVD EUA outbreaks (H1N1 through mpox) was removed, and a pointer to FDA's guidance "Emergency Use Authorization of Medical Products and Related Authorities" was added.
Templates. The general IVD EUA template is now described as reflecting FDA's current thinking on "data and information that should be submitted" (the draft also said "validation study recommendations"). FDA "may" (not "plans to") provide updated templates in a specific outbreak.
Pre-EUA described more fully. The final adds a description of the pre-EUA as a submission for early interaction and points to Section III.C of the EUA guidance.
2. Clinical Performance Evaluation (Section V.A)
30 positive / 30 negative is no longer a universal floor. The draft said a clinical evaluation "with at least 30 positive and 30 negative specimens" should be performed. The final reframes this as an *example*: for a viral nucleic acid test using established high-sensitivity technology, at an outbreak stage where samples and comparators are available, 30/30 is "typically adequate." The supporting footnote (prior emergencies used 50/50 contrived; the 12-sample COVID test that dropped to 71% PPA) is retained, so FDA still signals that fewer samples are risky.
Comparator definition softened. A highly sensitive comparator is "often" (not "typically") molecular, and the PPA ≥ 95% criterion is now given as an example of high sensitivity. The draft's specific recommendation that multi-analyte tests use an FDA-cleared/approved/authorized molecular comparator with prospective data from the past 5 years for non-emergency analytes was removed.
Hierarchy of alternative specimens reversed. In the draft, contrived specimens were presented first, with archived samples as an alternative. In the final, archived clinical specimens are the first fallback when natural prospective specimens aren't available, and contrived samples are used "when prospective and archived samples are not available." Archived specimen documentation should now also include specimen storage conditions.
Contrived sample numbers removed. The draft's minimum of 30 contrived positives (including at least 20 within 2-fold of the LoD) is gone.
Synthetic material language strengthened. The draft said synthetic material "could be considered"; the final says FDA "would likely consider" it acceptable when pathogen stocks are unavailable.
Condition of Authorization language. A CoA for a post-authorization clinical evaluation with natural specimens "would likely" be included (draft: "will typically").
Low-positive enrichment targets removed. The draft's COVID-era expectations (about 20% low positives overall, about 25% for molecular and 10–20% for antigen, with low positives defined as within 3 Ct of the comparator's LoD Ct) are gone. The final keeps only the general principle that non-prospective evaluations should represent the range of pathogen levels expected in the intended use population.
Study design specifics removed. Removed: sizing the study by prevalence to reach 30/30; the example of 30 sputum or 15 NP + 15 sputum; and the asymptomatic expectation of 20 positives / 100 negatives after a symptomatic claim. Enrichment strategies for asymptomatic claims remain possible but should now "minimize potential biases," and FDA consultation beforehand is still recommended.
New statistical expectation: Wilson Score confidence intervals. Clinical data analysis should now report PPA and NPA with corresponding 95% confidence intervals using the Wilson Score method. This is a new, concrete requirement industry should build into statistical analysis plans.
Numeric acceptance criteria removed. The draft's benchmarks (≥ 95% PPA/NPA for molecular tests; ≥ 80% PPA / ≥ 95% NPA for antigen tests; ≥ 98% NPA for asymptomatic screening; lower PPA acceptable for POC/home with mitigations) were removed. The final says targets depend on test type, indications, and benefit-risk, and that lower PPAs may be acceptable with mitigations such as "presumptive" negative reporting. The explicit reference to serial testing as a mitigation was dropped. Expect acceptance criteria to be set case by case with FDA.
Symptom onset analysis now more outcome-oriented. Data analysis should "account for" performance shifts relative to time from symptom onset "to support adequate clinical performance."
Human subject protection. New sentence: use of leftover, de-identified specimens under the informed consent policy does not obviate the other recommendations in the guidance.
3. Analytical Validation (Section V.B)
Limit of Detection
Swab and transport-media procedures removed. The draft's detailed instructions (spike swabs before immersing in VTM; apply contrived specimen directly to dry swabs; establish separate LoDs for VTM and dry processing) were dropped. Note that the clinical and specimen stability sections still treat different transport methods as distinct specimen types.
Stock material. The final calls for "well characterized, quantified natural stocks" and drops the explicit option of using a quantified known-positive clinical specimen (determined by an FDA-authorized test) to create dilutions.
New biosecurity reference for synthetic material. A new footnote directs manufacturers using synthetic nucleic acids to HHS/ASPR guidance, currently the 2023 HHS Screening Framework Guidance for Providers and Users of Synthetic Nucleic Acids. Manufacturers sourcing synthetic material should confirm their providers follow this framework.
Dilution series language softened. The preliminary LoD approach (2–3-fold series, 3 replicates) is now introduced with "Generally"; the 20-replicate confirmation, 95% criterion, and 3-fold step-down when 100% positivity is achieved are retained.
Removed: The COVID example that synthetic DNA/short RNA fragments overestimated performance; the explicit statement that CLSI EP17-A2 is FDA-recognized (still cited in a footnote); the restriction on simulated matrix and recombinant antigen moved to a footnote but is still in place.
Inclusivity
Core expectations retained: 100% sequence identity target, risk assessment for mismatches, in silico analysis for molecular tests, wet testing for antigen tests, monthly monitoring, aggregate impact not dropping performance below clinical recommendations, and risk analysis if sensitivity drops by more than 3-fold.
Changes: Wet testing of circulating variants for molecular tests is now qualified with "when available"; "past and present" variants language and its footnote were removed; the detailed workflow for mutations critical to test design (clinical specimen evaluation plus LoD wet testing) was removed; and the analysis should address impact on both clinical and analytical sensitivity. FDA "might" request additional testing based on its own monitoring.
Specimen Stability
CDC role clarified. The draft said acceptable conditions are "typically required" by CDC; the final says CDC "often communicates" them, and makes clear that a stability study is needed for conditions beyond CDC's specifications or where CDC has none.
Replicate numbers removed. The draft's 30 replicates at < 2-fold LoD and 10 at < 5-fold LoD are replaced by "around the LoD." The minimum of 10 negative specimens remains.
Language shifts from "labeling" to "proposed" conditions, and the home collection stability simulation guidance moved into a footnote.
Basic Safety, Software, Cybersecurity, and EMC (consolidated)
Four draft sections merged into one shorter section. The detailed EMC standards recommendations (IEC 61326-1, IEC 61326-2-6, ANSI/AAMI/IEC 60601-1-2 test levels, and acceptance criteria) were removed; the final simply says electrically powered components should demonstrate adequate EMC, pointing to FDA's EMC guidance. IEC 60601-1 for basic safety moved to a footnote.
Batteries are newly called out. Basic safety should be addressed for instrumentation "and/or batteries" not previously cleared, approved, or authorized for use with the test.
Software and cybersecurity scope reframed around instruments and device software functions used with the test. The referenced software guidance is now titled "Policy for Device Software Functions and Mobile Medical Applications."
Usability, User Comprehension, and Flex Studies (consolidated)
Usability now explicitly applies to POC tests, not just home collection and home use.
User comprehension is scoped to tests not involving a healthcare provider.
Combining studies is easier. The final says usability, user comprehension, and clinical performance studies may be combined. The draft recommended contacting FDA before combining usability with clinical evaluation.
Flex study details removed. The draft's list of conditions (read times 4x below to 3x above; specimen and buffer volumes 2x below/above; elution/mixing extremes; 40°C/95% RH and 5°C/5% RH; lighting; disturbance; orientation), testing to point of failure, 3 replicates per condition, and failure definitions are gone. The final keeps the hazard risk analysis first, then flex testing to identify the maximum deviation that still gives accurate results, with a pointer to the CLIA waiver guidance. Many manufacturers will likely still use the draft list as a practical benchmark.
New statement that recommendations may be adjusted given outbreak circumstances and "the important role of POC tests and home testing options."
Studies Demoted to "Potential Additional Testing"
These were stand-alone recommended studies in the draft and are now situational:
Cross-reactivity, microbial interference, and endogenous/exogenous interference are generally expected only if the clinical evaluation does not use real clinical specimens, or where FDA has accuracy concerns (e.g., novel approaches or performance signals). The draft's detailed design elements are gone, including the ≥ 80% in silico homology trigger for wet testing, 10^6 CFU/mL and 10^5 PFU/TCID50/mL concentrations, 3 + 10 replicate scheme, pooling rules, the recommended organism lists for respiratory and oral specimens, the interfering substance lists, CLSI EP07/EP37 references, and the warning against using VTM with lateral flow immunoassays. This is a significant reduction in default burden for tests validated with natural clinical specimens.
High-dose hook effect is still expected generally for antigen tests, but the 3–5 replicates per concentration design detail is gone.
Carry-over/cross-contamination now applies to "tests at risk" (e.g., automated liquid handling) rather than all workflows; the 5-run, 8 high positive / 8 negative checkerboard design was removed.
Fresh vs. frozen is still expected for tests using or validated with frozen specimens, but the 50-specimen minimum and ≥ 95% positive agreement criterion were removed.
Analytical equivalency is now something FDA "may recommend" to reduce the likelihood of additional clinical evaluation; the 3-fold LoD equivalence criterion and the instruction to proceed with the least sensitive configuration were removed.
Removed Entirely
Reagent stability no longer has a section. The draft's recommendations on unopened, shipping, open-kit, on-board, and freeze-thaw stability, CLSI EP25-A, and the option to rely temporarily on accelerated stability for a six-month shelf life with an FDA-agreed real-time study are all gone. Manufacturers should still expect to support labeled shelf life and storage claims, but the guidance no longer spells out an approach.
Section V general language about validating on initial production units, lots, or batches was removed. The final says validation "ideally" uses the final design and proposed instructions for use, and newly states that validation protocols and reports should be provided (objectives, procedures, and information on instruments, reagents, software, and other components).
Lawfully marketed components now explicitly include reagents as well as instruments, with an example that instrument-specific validation may not be needed for a test run on a lawfully marketed instrument. The right-of-reference mention in that paragraph was removed.
4. Change Protocols (formerly PCCPs)
Terminology change. FDA now uses "change protocols" and explains in a footnote that this distinguishes them from PCCPs, which apply only to marketing submissions. Industry should update internal SOPs and submission language accordingly.
New expectations. Change protocols should be reviewed by FDA and included in the initial authorization; should specify any labeling changes that would follow successful validation; and implementation requirements will generally be specified in a Condition of Authorization.
Examples updated. Adding new instruments or collection devices (the draft's example of extending shelf life/expiration date was dropped).
5. Additional Considerations for Certain Test Types (Section VI)
Multi-Analyte Panels
Largely unchanged in substance, condensed. The validation factors list is retained.
Home Specimen Collection and Home Use Tests (merged)
Two sections combined into one, with clear definitions: home collection means collecting and sending a specimen to a lab; home use means collecting and testing at home.
Home collection kit validation should evaluate the entire workflow, from kit distribution through home collection, storage, and shipping to a CLIA-certified lab.
Hazardous materials language softened. The draft said such kits were "generally not appropriate" unless they had specific safety features; the final says they "may not be appropriate without adequate risk mitigations." The PPE recommendation was removed.
Internal control exception removed. The draft noted an internal control may not be needed for specimen types shown appropriate for self-collection (e.g., anterior nasal swabs); that carve-out is gone. Mitigations now listed are video observation and/or an internal control.
Comparator specimen collection should follow comparator device labeling (e.g., by a healthcare provider), rather than always by a healthcare provider.
Software applications that interpret "or otherwise impact" test results should be evaluated in analytical and clinical studies (broader than "interpret ... or otherwise function as part of the test system").
Each proposed patient population (self-testing adults, supervised children, caregiver-tested children) should be evaluated following the guidance's validation recommendations.
Point-of-Care Tests
POC defined as professional near-patient settings.
The 1–2 site, 4–6 untrained operator design is now tied to CLIA Certificate of Waiver settings. The draft applied it to all POC clinical evaluations; the final specifies it for POC tests intended for use in waived settings.
Near-LoD study clarified. "Near the LoD" is now defined (e.g., ≤ 3-fold of the LoD), and testing is by the intended POC users. The 95% positive and negative agreement criterion remains.
Operator instruction. Operators should rely only on the QRI and receive no additional training "or intervention." The draft's explicit exclusion of supplemental materials (videos, apps) as a worst-case design was removed, as was "immediately" for real-time testing.
The differences are generated by AI models and can contain mistakes. Please double-check the content.